Detection of Phosphorus
|
Equipment & Reagents:
- Fume Hood
- Nebulizer - 250 ml, 20/40 Joint
Mfg - Kontes
Catalog # - K4225300250
Mfg - Chemglass
Catalog # - NC9755621
- Stirring hot plate
- Magnetic stir bars
- Assorted glassware
- Molybdenum VI Oxide - Aldrich Catalog Number - 22,181-3
- Molybdenum metal - Aldrich Catalog Number - 20,382-3
- Sulfuric acid
- Deionized water
Procedure:
Preparation of Reagent 1
- Dissolve 40.1g of molybdic anhydride in 1L of conc. sulfuric
acid with heat and stirring overnight.
- Cover the beaker or flask with plastic film (Saran wrap)
and secure with rubber band to prevent uptake of moisture.
- Produces a clear, light blue solution.
Preparation of Reagent 2
- Dissolve 1.78g of molybdenum metal using 500mL of Reagent
1 with heat and stirring overnight.
- Cover the beaker or flask with plastic film (Saran wrap)
and secure with rubber band to prevent uptake of moisture.
- Produces a dark blue solution.
Preparation of Phosphorus Spray
- To 120mL of deionized water add with stirring: 40mL Reagent
1 40mL Reagent 2
- Use extreme caution while adding reagents to water.
- Important! Always add reagents to water, not water to
reagents. Adding water to sulfuric acid solutions can be
explosive.
- Thoroughly mix the solution, allow to cool, and pour
into a nebulizer.
Application of Phosphorus Spray
- In a fume hood, spray TLC plates with molybdate solution
until plate is covered with reagent.
- Heat plate slightly until spots completely develop.
- Phosphorus containing compounds will appear as blue spots
on white background. Additional heating will char unsaturated
compounds.
Reference
- Ellingson, J.S. & William E.M. Lands (1968) "Phospholipid
Reactivation of Plasmalogen Metabolism", Lipids,
pp. 111-120.
Equipment & Reagents:
- Nebulizer
- Assorted glassware
- n-Butanol
- Ninhydrin reagent - Acros Catalog Number - 41572-0250
- Deionized water
Procedure:
Preparation of Ninhydrin Spray
- Saturate n-butanol with water by shaking 200mL of n-butanol
with 200mL of deionized water in a separatory funnel. Discard
the water phase (lower phase).
- Dissolve 200mg of ninhydrin reagent in saturated n-butanol
with stirring.
- Produces a clear, light yellow solution.
- Thoroughly mix the solution and pour into a nebulizer.
Application of Ninhydrin Spray
- In a fume hood, spray TLC plates with ninhydrin solution
until plate is covered with reagent.
- Heat plate until spots completely develop.
- Primary amine containing compounds will appear as pink
spots on white background.
Reference
- Skipski, V.P., Peterson, R.F., & M. Barclay (1962)
J. Lipid Res. 3:467.
|